5 set upol - titration parameters, Titration parameters – Metrohm tiamo 2.2 Manual User Manual
Page 740

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
726
■■■■■■■■
tiamo 2.2
Titrino
Range
0 ... 999 s
Default value
10 s
Selection
inf.
Stop time
The titration is stopped if the stop time defined here has elapsed since the
start of the titration. This parameter is only shown for Stop criterion =
time and Delay time = inf.
Titrino
Range
0 ... 999999 s
Selection
off
Default value
off
Endpoint 2
Note
Is activated only for EP1 with ≠ off.
on | off (Default value: off)
Check box for switching endpoint 2 on/off. If this is deactivated, then the
associated parameters will not be displayed.
EP2 with
Measured value for the second endpoint.
Range
-200.0 ... 200.0 µA
Selection
off
Default value
off
For additional parameters for Endpoint 2, see Endpoint 1.
5.6.3.4.6.5
SET Upol - Titration parameters
Tab: Method
▶ SET Upol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
Parameters defining the run of the titration.