6 Configuration
tiamo 2.2
■■■■■■■■
1311
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-30 mV
Upper limit
Upper intervention limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
30 mV
Warning limits for cell constant (only for conductivity sensors)
on | off (Default value: off)
If this option is enabled then the values for the cell constant will be shown
in orange if the limits are exceeded.
Lower limit
Lower warning limit for the cell constant.
Range
0.001 ... 500 /cm
Default value
2.0 /cm
Upper limit
Upper warning limit for the cell constant.
Range
0.001 ... 500 /cm
Default value
400 /cm
Intervention limits for cell constant (only for conductivity sen-
sors)
on | off (Default value: off)
If this option is enabled then the values for the cell constant will be shown
in red if the limits are exceeded.
Lower limit
Lower intervention limit for the cell constant.
Range
0.001 ... 500 /cm
Default value
0.001 /cm
Upper limit
Upper intervention limit for the cell constant.
Range
0.001 ... 500 /cm
Default value
500 /cm