Metrohm tiamo 2.2 Manual User Manual
Page 630

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
616
■■■■■■■■
tiamo 2.2
[Delete]
Deletes the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and temperature
for the minimum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
1.0 ... 2000.0 mV/mL
Default value
25.0 mV/mL
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and temperature
for the maximum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
1.0 ... 2000.0 mV/mL
Default value
25.0 mV/mL
Break-point evaluation
Note
This evaluation method is possible only with Titrando and 855.
Break-point evaluation
on | off (Default value: off)
A break-point evaluation is used to determine sharp changes of direction in
the titration curve.