Metrohm tiamo 2.2 Manual User Manual
Page 835

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
821
off
The measured vales will not be applied until after the maximum waiting
time has passed.
Min. waiting time
The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.
Range
0 ... 999999 s
Default value
1 s
Pause 1
Pause 1
Waiting time, e.g. until the electrode has settled down, before a start volume
is added.
Titrando, 855
Range
0 ... 999999 s
Default value
0 s
Start volume
Start volume
Volume to be added before the start of the titration at the dosing rate indi-
cated.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Titrino
Range
0.00 ... 999.99 mL
Default value
0.00 mL