5 Method

tiamo 2.2

■■■■■■■■

933

Range

2 ... 20

Default value

5

Window

A range (window) can be defined on the measured value axis or on the time
axis. The break point evaluation will only be carried out in the defined win-
dow. Only the first break point in the defined window will be recognized.

Selection

Measured value | time | off

Default value

off

Lower limit

Value for the lower limit of the window.

Time for the lower limit of the window.

Window = Measured value
Range

-200.0 ... 200.0 µA

Default value

-200.0 µA

Window = Time
Range

0 ... 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Value for the upper limit of the window.

Window = Measured value
Range

-200.0 ... 200.0 µA

Default value

200.0 µA

Window = Time
Range

1 ... 999999 s

Default value

999999 s

5.6.4.6.5

MEAS Upol - Additional measured values

Tab: Method

MEAS Upol Properties... Additional measured values

Command name

Name of the command.

Input

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in the
default settings in additional measured value columns Calculated # or
External # in the measuring point list .