8 Manual Control
tiamo 2.2
■■■■■■■■
1477
Range
1 ... 999 min
Selection
off
Default value
off
[Start]
Start heater. In the status display the current temperature is shown.
[Stop]
Stop heater.
Gas flow
Inlet
Select the inlet of the carrier gas.
Selection
Pump | Valve
Default value
Pump
Pump
Use of ambient air as carrier gas.
Valve
Use of gas out of a compressed gas cylinder.
Flow rate
874
Range
10 ... 150 mL/min
Default value
50 mL/min
Gas type
Gas type of the carrier gas.
Selection
Air | Nitrogen | Other gas
Default value
Air
Gas flow factor
Factor for the correct determination of the gas flow. This parameter is only
displayed for Gas type = Other gas. The values for frequently used carrier
gases are listed in the manual of the 774 and 874 Oven Sample Processor.
Range
1 ... 9.999
Default value
1.000
[Start]
Switch on gas flow. In the status display the running time since start is
shown.
[Stop]
Switch off gas flow.