5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1064

■■■■■■■■

tiamo 2.2

Tandem dosing

on | off (Default value: off)

If this option is enabled, the dosing is carried out without interruption with
a combination of two dosing devices so that the second dosing device is
dosing while the first one is being filled and vice-versa. This parameter is
displayed only for Titrandos, 846, 856, 867, 814, 815, 855, 864 and 874.

Dosing device

Selection of the second dosing device (exchange or dosing unit) to be used
for the dosing if the first dosing device is not available.

Titrando, 846, 856, 867

Selection

1 | 2 | 3 | 4

Default value

2

814, 815, 855, 864, 874

Selection

1 | 2 | 3

Default value

2

Solution

Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked.

Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Filling rate

Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during dosing
to the largest possible value.

Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min