Metrohm tiamo 2.2 Manual User Manual
Page 1145

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
1131
Stirrer type
Display of the stirrer type. Unknown is displayed either with devices unable
to read out the stirrer type or if no stirrer is connected
Selection
801 Magnetic Stirrer | 802 Rod Stirrer | 803 Ti
Stand magnetic stirrer | 804 Ti Stand rod stir-
rer | unknown
Stirring rate
Setting the stirring rate or rotational speed. The algebraic sign of the stirring
rate changes the direction in which the stirring is done. The default setting
8 corresponds to a rotational speed of 1000 rpm.
Titrando, 814, 815, 846, 855, 856, 867, 864, 874
Range
-15 ... 15
Default value
8
730, 778, 789
Range
0 ... 15
Default value
8
Action
Selection
Switch on | Switch off | Period of operation
Default value
Switch on
Switch on
Switch on the stirrer. The stirrer keeps switched on after the command
has finished.
Switch off
Switch off the stirrer.