background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2

■■■■■■■■

1189

5.6.10.4.2

LOOP - Properties

Dialog window: Method

LOOP Properties... LOOP - Loop #

Command name

Name of the command.

Input

25 characters

Stop criteria
Definition of three possible stop criteria. The first stop criterion to be fulfilled
will stop the loop.

Maximum run number

on | off (Default value: on)

If this option is enabled, then the maximum number of runs will be used as
stop criterion by repeat and while loops. If the maximum number of runs is
reached, the loop will be canceled at the time the LOOP command is started
again.

Range

0 ... 999

Default value

1

Maximum run time

on | off (Default value: off)

If this option is enabled, then the maximum run time will be used as stop
criterion by repeat and while loops. The starting point for the run time is the
first start of the loop command, which means that the run time for While
loops is started only after the loop has been run through once. If the max-
imum run time is reached, the loop will be canceled at the time the LOOP
command is started again.

Range

0 ... 999.9

Selection

min | s

Default value

min

Note

The run time is paused if the method is stopped with [HOLD].

Condition

on | off (Default value: off)

If this option is enabled, then the condition defined here will be used as stop
criterion by repeat and while loops.