Metrohm tiamo 2.2 Manual User Manual
Page 657

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
643
Smoothing factor
The higher the smoothing factor, the fewer endpoints will be found.
Range
2 ... 20
Default value
5
Window
A range (window) can be defined on the measured value axis, on the volume
axis or on the time axis. The break-point evaluation will only be carried out
in the defined window. Only the first break-point in the defined window will
be recognized.
Selection
Measured value | Volume | Time | off
Default value
off
Lower limit
Value for the lower limit of the window.
Window = Measured value
Range
-200.0 ... 200.0 µA
Default value
-200.0 µA
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Window = Measured value
Range
-200.0 ... 200.0 µA
Default value
200.0 µA
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
Window = Time
Range
0 ... 999999 s
Default value
999999 s