10 set upol - fix endpoint evaluation, 5 kft, 1 kft - overview – Metrohm tiamo 2.2 Manual User Manual

Page 748

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

734

■■■■■■■■

tiamo 2.2

[Properties]

Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880)
.

[Delete]

Deletes the external measured value selected in the table.

5.6.3.4.6.10

SET Upol - Fix endpoint evaluation

Dialog window: Method

SET Upol Properties... Additional evaluations Fix

endpoint evaluation

[New]/[Properties] Fix endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-200.0 ... 200.0 µA

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.5

KFT

5.6.3.5.1

KFT - Overview

Menu item: Method

Paste New command...

Command for volumetric Karl Fischer Titrations.

Principle
With this command, volumetric titrations for determining the water content
by the Karl Fischer method with automatic conditioning are carried out. A
typical feature of the Karl Fischer titration is the controlled reagent addition,
which depends on the control difference (current measured value – end-
point), with the aim of reaching the measured value given as being the
predefined target endpoint as quickly and accurately as possible. Overtitrat-
ing is avoided to a great extent. The titration termination at the endpoint
takes place either drift-controlled or after a waiting period. The volume
added up to the endpoint gives the calculable reagent consumption.

Advertising