5 Method
tiamo 2.2
■■■■■■■■
847
Titrino
Range
0.00 ... 999.99 mL
Default value
0.00 mL
Dosing rate
Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Pause 2
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the
end of all the start conditions.
Range
0 ... 999999 s
Default value
0 s
5.6.3.8.7.4
STAT U - Control parameters
Tab: Method
▶ STAT U ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Input
25 characters
Control point
Control point at
Definition of the current value to be controlled.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off