7 properties - glp – Metrohm tiamo 2.2 Manual User Manual

Page 1447

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Devices

tiamo 2.2

■■■■■■■■

1433

With the 774 Oven Sample Processor display of the initial temperature set
on the device only.
Range

50 ... 250 °C

Default value

off °C

Max. temperature

With the 774 Oven Sample Processor display of the maximum temperature
set on the device only.

With the 774 Oven Sample Processor display only.
Range

50 ... 250 °C

Default value

250 °C

Temperature correction

Enter the temperature correction. This parameter directly affects the tem-
perature control and allows it to be adjusted. In this way any temperature
difference that may occur between the oven and sample temperature can
be compensated.

With the 774 Oven Sample Processor display of the temperature correction
set on the device only.
Range

-10 ... +10 °C

Default value

0 °C

7.14.7

Properties - GLP

Tab: Configuration

Device [Edit] Properties... 'Properties - 'Device type' -

'Device name'

GLP

GLP test date

Date of the last GLP test. This date can be selected by clicking on

in the

dialog window Select date (see Chapter 2.5.1, page 83).

Comment GLP test

Comment on GLP test.

Input

1000 characters

Monitoring of GLP validity

Monitoring of GLP validity

on | off (Default value: off)

If this option is switched on then the time interval for the GLP test will be
monitored.