6 cal loop conc, 1 cal loop conc - overview, Cal loop conc – Metrohm tiamo 2.2 Manual User Manual
Page 1054

5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1040
■■■■■■■■
tiamo 2.2
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
5.6.5.6
CAL LOOP Conc
5.6.5.6.1
CAL LOOP Conc - Overview
Dialog window: Method
▶ CAL LOOP Conc ▶ Properties... ▶ CAL LOOP Conc - 'Com-
mand name'
Loop command for the calibration of ion-selective electrodes, which is
used together with the command CAL MEAS Conc.
Appearance
The command has the following appearance:
Parameters
The parameters for the command CAL LOOP Conc are configured in the
following dialog window:
■
Principle
For calibrating ion-selective electrodes (ISE), a linear correlation of the meas-
ured potential U
i
and the logarithm of the sought ion concentration c
i
cannot
always be assumed. In ion measurements, interfering ions have an influence
on the measurement and that is why this effect is perceptible by a flattening
of the curve, especially in low concentration ranges.