5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1096
■■■■■■■■
tiamo 2.2
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Range
0.1 ... 999999 s
Default value
5.0 s
Titrino
Range
1 ... 999999 s
Default value
5 s
5.6.6.3.5.4
DOS U - Stop conditions
Tab: Method
▶ DOS U ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
The conditions for stopping the dosing are defined on this tab. If more than
one stop condition is in effect at the same time, then the criterion which is
fulfilled first will stop the dosing.
Stop volume
Stops when the given volume has been added after the start of the dosing.
The stop volume should be adapted to suit the measuring vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100.000 mL
Selection
off
Titrino
Range
0 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Range
0 ... 999999 s
Selection
off
Default value
off