Metrohm tiamo 2.2 Manual User Manual

Page 1470

Advertising
background image

7.18 Barcode reader

■■■■■■■■■■■■■■■■■■■■■■

1456

■■■■■■■■

tiamo 2.2

Input

1000 characters

Monitoring of GLP validity

Monitoring of GLP validity

on | off (Default value: off)

If this option is switched on then the time interval for the GLP test will be
monitored.

Monitoring can only be switched on when a date has been entered in the
field GLP test date.

GLP test interval

Time interval to next GLP test. If a value is entered here then the date in the
field Next GLP test will be adapted automatically.

Only editable when the monitoring is activated.
Range

1 ... 999 Days

Default value

999 Days

Next GLP test

Date on which the next GLP test is to be carried out. The date can be selected

by clicking on the symbol

in the dialog window Select date (see Chapter

2.5.1, page 83). After the date has been entered, the field GLP test inter-
val
will be adapted automatically.

Only editable when the monitoring is activated.
Default value

    GLP test date + 999 days

Message
The message options can only be edited if monitoring is switched on.

Message by E-mail

on | off (Default value: off)

The message is additionally sent to the e-mail address defined under [E-
mail]
if this option is activated. The message is sent in text format.

[E-mail]

[E-Mail] opens the Send e-mail window (see Chapter 2.6.1, page 86).

Acoustic signal

on | off (Default value: off)

If this option is enabled, an acoustic signal will be emitted additionally to the
message defined above

Advertising