5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
666
■■■■■■■■
tiamo 2.2
Titrando, 855
Range
0 ... 999999 s
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration. The
maximum filling rate depends on the cylinder volume of the exchange unit
or dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.4.3.7
SET pH - Conditioning
Tab: Method
▶ SET pH ▶ Properties... ▶ Conditioning
Command name
Name of the command.
Input
25 characters
Conditioning
on | off (Default value: off)
Enable/Disable the conditioning. If the conditioning is disabled, the following
parameters are not visible.
Start drift
The determination can only be started when the currently measured drift is
less than the entered start drift entered here. If this is the case, then the
status displayed in the Run subwindow in the Workplace program part
changes to COND READY.
Titrando, 855
Range
1 ... 999 µL/min
Default value
20 µL/min