3 add - dosing parameters – Metrohm tiamo 2.2 Manual User Manual
Page 1079

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
1065
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the dosing has finished. This parameter is displayed only for Titrandos, 846,
856, 867, 814, 815, 855, 864 and 874.
5.6.6.2.3
ADD - Dosing parameters
Tab: Method
▶ ADD ▶ Properties... ▶ Dosing parameters
Command name
Name of the command.
Input
25 characters
The volume to be added and the dosing and filling rate are defined on this
tab.
Volume
Fixed volume that is to be dosed.
Titrando, 846, 814, 815, 855, 864, 874
Range
0.00000 ... 99999.9 mL
Default value
10.0000 mL
Titrino
Range
0.0001 ... 9999.0 mL
Default value
10.0000 mL