5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
730
■■■■■■■■
tiamo 2.2
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 99.9 µL/min
Default value
0.0 µL/min
Stop volume
Stops when the given volume has been added after the start of the titration.
The stop volume should be adapted to the size of the titration cell in order
to prevent any overflow.
Titrando, 855
Range
0.000000 ... 9999.99 mL
Default value
20.00 mL
Selection
off
Default value
off
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Titrando, 855
Range
1 ... 999999 s
Selection
off
Default value
off
Stabilizing time
Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.
Range
0 ... 999 s
Default value
0 s
Sample addition time
Waiting time between stop of conditioning and start of titration during
which the sample can be added.
Range
0 ... 99 s
Default value
0 s