6 kft upol - stop conditions, Stop conditions – Metrohm tiamo 2.2 Manual User Manual

Page 784

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

770

■■■■■■■■

tiamo 2.2

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando
Range

0.1 ... 999999 s

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s

Default value

2.00 s

5.6.3.5.3.6

KFT Upol - Stop conditions

Tab: Method

KFT Upol Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.

Stop volume

Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.

Titrando
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.00 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Advertising