7 met ph - additional evaluations – Metrohm tiamo 2.2 Manual User Manual

Page 577

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2

■■■■■■■■

563

ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.

Evaluation with measured value window (pH)

[New]

Opens the dialog window Measured value # in which the parameters for
a new window can be entered (see Chapter 5.6.3.3.2.9, page 569).

[Properties]

Opens the dialog window Measured value # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.3.2.9, page 569).

[Delete]

Deletes the window selected in the table.

Evaluation with volume window (mL)

[New]

Opens the dialog window Volume window # in which the parameters for
a new window can be entered (see Chapter 5.6.3.3.2.10, page 570).

[Properties]

Opens the dialog window Volume window # in which the parameters for
the selected window can be edited (see Chapter 5.6.3.3.2.10, page
570)
.

[Delete]

Deletes the window selected in the table.

5.6.3.3.2.7

MET pH - Additional evaluations

Tab: Method

▶ MET pH ▶ Properties... ▶ Additional evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of titration curves can be
activated and defined on this tab: