7 kft upol - conditioning, Conditioning – Metrohm tiamo 2.2 Manual User Manual

Page 785

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.2

■■■■■■■■

771

Stop time

Titrando
Range

0 ... 999999 s

Selection

off

Default value

off

off
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.

Filling rate

Rate at which the dosing cylinder is to be filled after titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
device used.

Titrando
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.5.3.7

KFT Upol - Conditioning

Tab: Method

KFT Upol Properties... Conditioning

Command name

Name of the command.

Input

25 characters

Conditioning

Conditioning

on | off (Default value: off)

Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.

Start drift

The determination can only be started when the currently measured drift is
less than the start drift entered here. In such cases, the Status in the
Run subwindow displayed on the Single determination or Determina-
tion series
tab will then switch to COND READY.