4 dos ph - stop conditions, Stop conditions – Metrohm tiamo 2.2 Manual User Manual
Page 1095

■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2
■■■■■■■■
1081
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando, 855
Range
0.1 ... 999999 s
Default value
5.0 s
Titrino
Range
1 ... 999999 s
Default value
5 s
5.6.6.3.4.4
DOS pH - Stop conditions
Tab: Method
▶ DOS pH ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
The conditions for stopping the dosing are defined on this tab. If more than
one stop condition is in effect at the same time, then the criterion which is
fulfilled first will stop the dosing.
Stop volume
Stops when the given volume has been added after the start of the dosing.
The stop volume should be adapted to suit the sample weight or the titration
vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100.000 mL
Selection
off
off
Means that no stop will take place.
Titrino
Range
0 ... 9999.99 mL
Default value
100.00 mL
Selection
off
off
Means that no stop will take place.