How do I recalibrate the

blower speed

filter life gauge

airflow sensor reading

of the Logic+?

1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.



4.

Press the

button until the

Calibration

option is highlighted (it

will turn white).

5.

Press

OK/Mute

to start the calibration procedure.





6.

You will see a

warning screen

alerting you that you are about to

alter the BSC’s settings.

7.

Press

OK/Mute

to input the password.




8.

When requested for the password press

Light

,

UV light

,

Timer

,

Timer

then

OK/Mute

.





9.

If the password is properly entered, the first certification screen

will display the blower motor’s PWM setting.

10.

Press the

or

button to increase or decrease the blower

speed as needed. When the speed is set properly, press the

OK/Mute

button to lock the new blower speed setting, and

continue to the filter life screen.

MAIN MENU

1. MyLogic

TM

2. Settings

3. Service

 

or

to choose,

OK to select, MENU to escape

or

to choose,

OK to select, MENU to return

Current Settings

Restore Default

Diagnostics

Configuration

Calibration

 

Service

ATTENTION!

The following screens are used

to alter the operation of the

Logic. They should only be used

by a qualified certifier as part

of the certification process.

Calibration

OK to proceed, MENU to return

PASSWORD:

Calibration

The ECM setting is: 80
The current RPM is: 1050
The average RPM is: 1020

Calibration

Motor Speed Adjustment

or

to change motor speed,

OK to select, MENU to finish

 