Labconco PURICARE Procedure Station Models 31260 User Manual

Page 81

Advertising
background image

Product Service 1-800-522-7658

81

1.

Turn on the photometer and allow it to operate for a minimum of 5 minutes.

Leave the valve in the “CLEAR” setting.

2.

Check the oil level of the generator – it must be within 1/8” of the level line

when the generator is sitting level.

3.

Establish the 100% and 0% concentration levels for the photometer.

4.

Remove the diffuser by raising the sash as high as it will go, and then removing

the two acorn nuts with a 7/16” wrench or socket, as shown in Figure 5-11.

Figure 5-11

Removing the diffuser acorn nuts (Dress

panel and lamps removed for clarity

only).


Note: The diffuser acorn nuts are secured to the studs with a removable thread locking

compound to prevent them from vibrating loose during shipment of the Logic+. There

may be some resistance the first time the nuts are removed; this is normal. No

additional thread locking compound need be applied to these acorn nuts unless the unit

is being prepared for shipment.

5.

Lower the front edge of the diffuser until it clears the bottom of the sash, and

then pull the diffuser straight out of the cabinet, as shown in Figure 5-12.

Figure 5-12


6.

The Logic+ is now ready for the Supply HEPA Filter Leak Test. Set the

photometer sampling valve to “DOWNSTREAM”. Ensure there is a proper vacuum

at the sampling port of the pistol.