Product Service 1-800-522-7658

152

G4 Paper Prefilter Kits

34060-00, -01,-02, -03

Screen Prefilter Kits

34061-00, -01,-02, -03

The Prefilter provides a convenient means to prevent large particles from being

pulled into the blower and HEPA filter of the biological safety cabinet. The
Prefilter is a simple addition and easily replaced.

Please note that this accessory was tested at NSF and validated to

have no adverse affect on containment.

Installation Procedure

Note: The work surface of the cabinet and the towel catch must be thoroughly

decontaminated before removing either.

1.

Remove the work surface by lifting the front edge with the knobs. Slide the work

surface forward to disengage

the rear tabs. Support the

underneath side with one

hand while stabilizing the

front edge with the other.

2.

Lift the front end of the center

support straight up to release

it from the locating pins (A).

Set the support aside in the
work area.

3.

Access the towel catch by

pivoting the bottom out

toward you (B). Surface

decontaminate the towel

catch before removing it from

the cabinet.