Section Seven –

Electrical System Service Operations

NOTE: In order to properly service the Logic+, you must obtain a Logic+

Certifier Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by

contacting Labconco Customer Service Department at 800-821-5525.

Fluorescent Lamp

Lamp Identification

All Logic+s use dual fluorescent lamps. The lamp information is shown in Table
8-1. The code of the lamp part number is as follows:

F25T8 TL741

F = fluorescent lamp

25 = the lamp Wattage

T8 = the lamp mounting configuration

TL741 = the phosphor used in the lamp

Note: The phosphor of the lamp determines the color and intensity of

the lamp. Not matching the phosphor may affect the quality and/or

color of the light.

Table 7-1

Fluorescent Lamp Specifications

Cabinet Width (feet)

3

4

5

6

Fluorescent Lamps (2 each)

F25T8
TL741

F32T8
TL741

F40T8

SP41

F40T8

SP41

Fluor. Lamp Labconco Part #

9721901

9721900

9721903

9721903

Fluor. Ballast Labconco Part #

3838100

3838100

3838100

3838100

Fluor. Ballast Robertson Part #

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

ISL332T

8MVW

Changing the Fluorescent Lamp

1.

Unplug the cabinet or turn the System Reset Switch, located on the top of the

cabinet, off.

2.

Remove the front dress panel as noted in Figure 6-1.

3.

Remove the fluorescent lamps by pulling the lamp sockets straight off each end of

the lamp, and releasing both lamps from the spring clips that secure them in place.

4.

Install the new lamps by reversing the removal procedure.