Type B HEPA Leak Test Information

All Logic+ models were tested to the HEPA Filter Leak Test as described in ANSI/NSF

Standard 49:2012.

Note: ensure all internal surfaces of the Logic+ have been appropriately

decontaminated before proceeding.

Type B Supply HEPA Filter Leak Test

All Logic+ models were tested to the HEPA Filter Leak Test as described in ANSI/NSF

Standard 49:2012.

Note: ensure all internal surfaces of the Logic+ have been appropriately

decontaminated before proceeding.

Note: For all Logic+ models, the prefilter element should be removed, and the

aerosol generator should be placed nearby to inject the aerosol as shown in

Figure 5-9.

Figure 5-9

Place aerosol generator on top of the electronics
module


Remove the prefilter before starting the test




Measuring the Upstream Concentration

The Logic+ utilizes a unique aerosol access tube for sampling the upstream

concentration of aerosol. If after reading these instructions, you have further questions,
please contact Labconco's Product Service Department.

1.

Remove the work surface by lifting the handles on either end. Pull the work

surface straight out of the Logic+.

2.

Locate the access tube on right side of the rear plenum wall, passing through the

towel catch. As shown in Figure 5-10.