Labconco PURICARE Procedure Station Models 31260 User Manual

Page 141

Advertising
background image

Product Service 1-800-522-7658

141

9.

The first screen allows you to set the Type of cabinet. The only difference

between the Type A1 and Type A2 configuration setting is that the airflow

velocity limits for normal operation of the optional airflow

sensor (because the Type A1 has a nominal inflow of 85

FPM, and the Type A2 is 105 FPM). If the air velocity goes

out of range, the alarm will be displayed, but the cabinet will

continue to run. Selecting Type B2 will cause an inflow

alarm if the inflow velocity drops below the alarm setpoint,

and will result in the cabinet blower stopping. Use the

or

buttons to select the cabinet Type, and then press

OK/Mute.

10.

The second screen allows you to configure the cabinet for a

canopy or not. Select the “a canopy” option.

11.

This screen configures the unit for a UV light or not. After

your selection is chosen, and you press

OK/Mute

, the unit

will shut off and then reset in the new configuration.


Configuration

Type A Canopy Option

This cabinet has:

or

to choose,

OK to select, MENU to escape

a canopy

no canopy

 

This Logic is configured:

or

to choose,

OK to select, MENU to return

to have a UV light

to NOT have a UV light

 

UV Settings