Quickstart, Main menu 1. mylogic, Settings 3. service – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 16: Service

Advertising
background image

Product Service 1-800-522-7658

16

Quickstart


How do I run a diagnostic on the keypad, relays and

contacts?

1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.





4.

Press the

button until the Diagnostic

Diagnostic

option is

highlighted (it will turn white).

5.

Press

OK/Mute

to enter the Diagnostic submenu.





6.

Press the

button until the Diagnostic

Keypad-Relays

option is

highlighted (it will turn white). Press

OK/Mute

.





7.

Press any button on the keypad (except

Menu

), and note the

display.







MAIN MENU

1. MyLogic

TM

2. Settings

3. Service

 

or

to choose,

OK to select, MENU to escape

or

to choose,

OK to select, MENU to return

Current Settings

Restore Default

Calibration
Configuration

Diagnostics

 

Service

or

to choose,

OK to select, MENU to escape

Keypad -Relays

Sash sensors

Canopy Alarm
Airflow Sensor

 

Diagnostics