9.
If the password is properly entered, the first certification screen will display the
blower motor’s PWM setting.
10.
Press the
▲
or
▼
button to increase or decrease the
blower speed as needed. When the speed is set properly,
press the
OK/Mute
button to lock the new blower speed
setting, and continue to the filter life screen.
11.
If this is an initial certification or a recertification after
HEPA filter replacement, select
Reset Filter Gauge to
100%
. If this is an annual recertification, select
Leave
Gauge unchanged
. If you want to set the filter gauge at a
preset value, select
Set Gauge to new setting
. Then press
OK/Mute
.
The Type A units with airflow sensor will ask you to now
calibrate the average downflow to match the value you
obtained when you measured it.
12.
If the unit is a Type A without an airflow sensor, after the
filter gauge option is selected and you press
OK/Mute
,
the Logic+ will power down, and then restart with the new
settings. If the Unit is a Type B2 or a Type A with an
airflow sensor, you will then be asked to calibrate the
inflow sensor to match the value you obtained when you
measured the inflow. . After the inflow is calibrated, and
you press
OK/Mute
, the Logic+ will power down, and
then restart with the new settings.
This will end the calibration mode, and the unit will automatically turn off and then
restart, displaying inflow and downflow.
The ECM setting is: 80
The current RPM is: 1050
The average RPM is: 1020
Calibration
Motor Speed Adjustment
or
to change motor speed,
OK to select, MENU to finish
Calibration
Filter Gauge Adjustment
or
to change filter gauge,
OK to select, MENU to finish
Reset Filter Gauge to 100%
Leave Gauge unchanged
Set Gauge to new setting
The Inflow displayed
value is: 102 FPM
Calibration
Inflow Display
or
to change display value,
UV Light to reset to 105 FPM,
OK to select, MENU to finish