Labconco PURICARE Procedure Station Models 31260 User Manual

Page 145

Advertising
background image

Product Service 1-800-522-7658

145

9.

If the password is properly entered, the first certification screen will display the

blower motor’s PWM setting.

10.

Press the

or

button to increase or decrease the

blower speed as needed. When the speed is set properly,

press the

OK/Mute

button to lock the new blower speed

setting, and continue to the filter life screen.


11.

If this is an initial certification or a recertification after

HEPA filter replacement, select

Reset Filter Gauge to

100%

. If this is an annual recertification, select

Leave

Gauge unchanged

. If you want to set the filter gauge at a

preset value, select

Set Gauge to new setting

. Then press

OK/Mute

.

The Type A units with airflow sensor will ask you to now

calibrate the average downflow to match the value you

obtained when you measured it.

12.

If the unit is a Type A without an airflow sensor, after the

filter gauge option is selected and you press

OK/Mute

,

the Logic+ will power down, and then restart with the new

settings. If the Unit is a Type B2 or a Type A with an

airflow sensor, you will then be asked to calibrate the

inflow sensor to match the value you obtained when you

measured the inflow. . After the inflow is calibrated, and

you press

OK/Mute

, the Logic+ will power down, and

then restart with the new settings.


This will end the calibration mode, and the unit will automatically turn off and then

restart, displaying inflow and downflow.

The ECM setting is: 80
The current RPM is: 1050
The average RPM is: 1020

Calibration

Motor Speed Adjustment

or

to change motor speed,

OK to select, MENU to finish

 

Calibration

Filter Gauge Adjustment

or

to change filter gauge,

OK to select, MENU to finish

 

Reset Filter Gauge to 100%

Leave Gauge unchanged

Set Gauge to new setting

The Inflow displayed
value is: 102 FPM

Calibration

Inflow Display

or

to change display value,

UV Light to reset to 105 FPM,
OK to select, MENU to finish

 