Vibration (optional)

The vibration level should be measured as described in ANSI/NSF Standard 49:2012.

Note: The vibration meter should have a range of 20-20,000 Hz, and have an

output in displacement.

1.

Vibration level should be recorded from a single point, on the geometric center

of the work surface.

2.

Take a background reading with the cabinet blower off at the test point.

3.

Start the cabinet blower, and take a vibration reading.

Acceptance

Vibration should be less than 0.0002 inches displacement when the cabinet is new.

UV intensity (optional)

Note: The UV radiometer should measure light at a wavelength of 254nm.

1.

UV intensity should be recorded from a single point, on the geometric center of

the work surface.

2.

Take a background reading with the cabinet blower off at the test point.

3.

Close the sash, turn on the UV light, and let it warm up for at least 5 minutes.

Acceptance

The UV intensity should be at least 200 uW-cm

2

.

Most common causes for insufficient UV light levels are:

d.

Malfunctioning lamp(s)

e.

Old lamp(s)

f.

Dusty/dirty lamp surfaces