Labconco PURICARE Procedure Station Models 31260 User Manual

Page 30

Advertising
background image

Product Service 1-800-522-7658

30

QuickChart – Logic+ PuriCare Procedure Stations
Imperial Units

Model

31241

31142

31261

31162

Type

A2

A1

A2

A1

Cabinet Size (in feet)

4

4

6

6

Sash Opening (inches)

10

12

10

12

Starting Serial #

1303_

1303_

1303_

1303_

Downflow Data

Nominal Avg. Downflow (FPM)

55+/-5

55+/-5

55+/-5

55+/-5

grid # of points (rows x columns)

24 (3x8)

24 (3x8)

36(3x12)

36(3x12)

Grid distance from back & sides

6.0

6.0

6.0

6.0

Distance between rows

5.55

5.55

5.55

5.55

Distance between columns

5.21

5.21

5.50

5.50

Inflow Data

Nominal Average Inflow (FPM)

105+/-5

85+/-5

105+/-5

85+/-5

Sash Open Area (Sq. Ft)

3.37

4.04

5.03

6.04

Nominal Avg. Exhaust Vol. (CFM)

354

344

529

514

Avg. Exhaust Vol. Range (CFM)

337-371

323-364

503-553

483-544

Secondary Inflow Data

3

Sash Opening Template

1

3.81

3.81

4.95

4.95

Sensor distance (inches)

2

3.25

3.25

4

4

# of Test points

8

8

12

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (FPM)

354

344

289

281

Avg. Inflow Vel. Range (FPM)

337-371

323-364

275-302

264-297

Correction Factor (CF)

1

1

1.83

1.83

Average Inflow Volume Range

(AIV) (Avg. velocity x CF)

337-371

323-364

503-553

483-544

Sash Open Area (Sq. Ft)

3.37

4.04

5.03

6.04

Inflow Velocity Range

(AIV / Sash open area)

100-110

85+/-5

100-110

85+/-5

HEPA Filter Leak Test Data

Air Displacement (CFM)

794

794

1189

1189

Laskin Nozzles needed

2

2

2

2

Theoretical aerosol conc. (ug/l)

3

34

34

23

23

Actual aerosol conc. (ug/l)

4

18

18

15

15

Supply HEPA Data

Width x Depth x Height (in.)

5

48x18x3.06

48x18x3.06

72x18x3.06

72x18x3.06

Performance (CFM)

445

445

665

665

Performance (Pressure in. H

2

O)

<.45

<.45

<.42

<.42

Labconco P/N

3838401

3838401

3838403

3838403

Exhaust HEPA Data

Width x Depth x Height (in.)

5

26x18x5.88

26x18x5.88

48x18x5.88

48x18x5.88

Performance (CFM)

353

353

529

529

Performance (Pressure in. H

2

O)

< 0.28”

< 0.28”

< 0.28”

< 0.28”

Labconco P/N

3838501

3838501

3838503

3838503

Motor/Blower Data

Labconco P/N

6

3832201

3832201

3832208

3832208

Motor HP

½

½

¾

¾

Fluorescent Lamp Data

Fluorescent Lamps (2 each)

F32T8
TL741

F32T8
TL741

F40T8

SP41

F40T8

SP41

Fluor. Lamp Labconco Part #

9721900

9721900

9721903

9721903

Fluor. Ballast Labconco Part #

3838100

3838100

3838100

3838100

Ventus Canopy Data

Labconco Canopy P/N

3889201

3889201

3889203

3889203

Canopy Slot Area (F

2

)

.12

.12

.12

.12

Canopy Inflow Range (CFM)

20-100

20-100

20-100

20-100

Nominal Canopy Slot Velocity (FPM)

240-260

240-260

240-260

240-260

Nominal Canopy Slot Volume (CFM)

50

50

50

50

Canopy Vacuum (In. H

2

O)

0.1-0.15

0.1-0.15

0.1-0.15

0.1-0.15

Advertising