7.

Using a 1/2-inch wrench, remove the two upper and one lower blower mounting

bolts, as shown in Figure 6-29.

Figure 6-29


Blower Mounting Bolts







The motor/blower assembly is heavy. Handle with care.

8.

While supporting the blower assembly, pull the assembly straight out the

plenum, as shown in Figure 6-30.

Figure 6-30


9.

To replace the motor/blower assembly, reverse the above procedure.

Note: During reinstallation of the blower assembly, angle the rear blower

bracket into the plenum blower opening. Make sure to engage the rear

blower bracket inside the plenum blower hole. After reinstalling and

tightening the three blower bolts, ensure that the rear blower bracket is

correctly engaged by trying to move the blower assembly. If it is rigidly

captured on the plenum, then the rear bracket is properly engaged. If the

rear of the blower assembly moves freely, then the blower must be

reinstalled.
10.

Continue the reassembly. Plug in the cabinet and recertify it before use.