136

Section 1. - Canopy Installation Procedure

1.

Turn off the Biosafety Cabinet.

2.

Locate and remove the four acorn nuts and

lockwashers that secure each side of the exhaust

cover and the angle bracket to the top of the unit.

3.

Lift the cover and angles straight up off of the top of

the cabinet. Keep the nuts and washers; you may

also want to retain the cover and angles for the

future, if the cabinet will be disconnected from the

exhaust system.


4.

Install the replacement brackets as shown,

with the lockwashers, and tighten the acorn

nuts.

5.

Carefully place the canopy between the

brackets. Ensure the canopy fits on the inside

of the exhaust filter flanges assembly as

shown. Line up the holes in the canopy

connection with the screw holes in the filter

flanges.

6.

Install the four bolts, washers and lockwashers supplied with the canopy

connection kit, and tighten them securely to the exhaust assembly as shown.