Circuit breaker, Resetting a circuit breaker, Circuit breaker specifications – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 125

Advertising
background image

Product Service 1-800-522-7658

125

Circuit Breaker

Resetting a circuit breaker

To reset any of the circuit breakers located on the left side of the electronics

module, as shown in Figure 7-4, depress the white button until it sets.

Figure 7-4


Circuit Breakers- The

front breaker protects

the electrical outlets,

the rear breaker
protects the motor

and lights.











Circuit Breaker Specifications

Circuit breaker specifications are shown in Table 7-2, below.

Table 7-2

Circuit Breaker Specifications, 100/115 VAC

Cabinet Width (feet)

3

4

5

6

Motor/Light Breaker Rating,

Amps

8

8

12

12

Motor/Light Breaker, Labconco

Part #

1327208 1327208

1327201 1327201

Motor/Light Breaker,

Mechanical Products Part #

1600-062-

080-007

1600-062-

080-007

1600-062-

120-007

1600-062-

120-007

Outlet Breaker Rating, Amps

5

5

5

5

Outlet Breaker, Labconco

Part #

1327204 1327204

1327204

1327204

Outlet Breaker, Mechanical

Products Part #

1600-062-

050-007

1600-062-

050-007

1600-062-

050-007

1600-062-

050-007

Advertising