Type B2 Prefilter Replacement

NOTE: The prefilter is not washable; it must be replaced when it becomes

dirty.
1.

Shut off the cabinet blower.

2.

Using an appropriate step or ladder, remove the two wing nuts that secure the

prefilter, located next to the electronics module, as shown in Figure 6-24.

3.

Remove and discard the prefilter, and replace it with a new one. Retighten the two

wing nuts that secure the prefilter.

Figure 6-24



Prefilter wing nuts


Type B2 HEPA Filter Replacement

CAUTION: The cabinet must be properly decontaminated before servicing the

exhaust HEPA filter. Following replacement of either HEPA filter, the cabinet

must be recertified.

CAUTION: HEPA filters can be awkward to handle and heavy. Use appropriate

lifting techniques to remove and handle it. Use two people to remove the

larger filters if possible.

1.

Unplug the cabinet.

2.

Remove the front dress panel as shown in Figure 6-1.

To Remove the Exhaust HEPA Filter:
3.

Loosen all of the Phillips screws on the exhaust filter cover.

4.

Loosen the four sealed exhaust filter bolts on the top of the cabinet hull, as

shown in Figure 6-25. DO NOT remove the bolts; only loosen them until the

exhaust HEPA filter is free of the filter frame.

5.

Pull the exhaust HEPA filter straight out of the cabinet.