Sash position switches, Sash closure position switch, Sash operating position switch – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 101

Advertising
background image

Product Service 1-800-522-7658

101

Sash Position Switches

Two microswitches are used to indicate the sash position. The switches are the same,

Labconco part # 3832400. A photo of both switches is shown in Figure 6-6.

Figure 6-6

Sash Closure Position Switch

Sash Operating Position Switch




Sash Closure Position Switch

The sash closure position switch signals the display microprocessor that the sash

is open or closed allowing proper operation of the Blower, fluorescent or UV

lights. It is wired in the normally open configuration. If the switch does not

actuate properly when the sash is within ¼-inch of being fully closed, ensure the

switch follower is properly actuating against the weight. See Figure 6-7.

Sash Operating Position Switch

The sash operating position switch signals the display microprocessor that the

sash is open too far for proper operation, as shown in Figure 6-8. It is wired in

the normally open configuration. If the switch does not actuate properly when

the sash is within ¼-inch of its operating position, ensure the switch follower is

properly actuating against the weight.