Page 35

Advertising
background image

Product Service 1-800-522-7658

35

QuickChart – Cell Logic+ Type A2 BSCs

footnotes

Logic Model Identification

The primary serial tag is on the lower outside edge of the right corner post.

The secondary serial tag is located on the front of the electronics module on

the top right side of the cabinet.

The first two digits of the serial number are the year of production; the next

two are the month. The next 5 digits are the sequence of production, and
the letter following the serial number is the revision level of the cabinet.

Downflow Test Specifications

All models are classified as uniform downflow.

All tests performed as described in ANSI/NSF Standard 49:2007.

UV Lamp, IV bar and all other accessories must be removed before

measuring downflow.

Inflow Test Specifications

All tests performed as described in ANSI/NSF Standard 49:2007.

Secondary Inflow Test Specifications

Must use Labconco holder P/N 3836405 to perform this test properly.

1.

Use the appropriate template included in Certifier Kit# 3858403.

2.

Measured from the bottom edge of the probe holder to the center of the

thermal anemometer sensor element.

* - Locate the single row of holes at the front of the grille. Mark the 6

th

hole

from the side wall and subsequent test points every 9 holes until the number of

test points marked equals the width of the cabinet in feet (3-foot cabinet, mark
the first 3 points). Repeat for the opposite side.

HEPA Filter Leak Test Specifications

3.

Based on mineral oil.

4.

The actual aerosol concentration is what was observed during testing.

Aerosol generator should be placed in the left rear corner of the work area, pointing

toward the rear grille. For uncontaminated units, the upstream concentration can be
sampled from the tube located under the work surface

Metric calculation of concentration is:

# of Laskin nozzles @ 138K Pascals x 6.372/Volume of air in m

3

/sec. =

Conc. In ug/l


Supply and Exhaust HEPA Filter Specifications

5.

Without gasket

Motor / Blower Specifications

6.

Each motor must be programmed by Labconco for the appropriate width

cabinet. The speed control settings will fluctuate depending on local
temperature and pressure.

UV lamp Specifications

For all models, the UV lamp number is G30T8. The Labconco part number is

1271300.

For all 115/230 VAC models, the ballast assembly is Labconco part number

3829901.

For all 100 VAC models, the ballast assembly is Labconco part number
3830600.