Quickstart – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 14

Advertising
background image

Product Service 1-800-522-7658

14

Quickstart


How do I change the Configuration of the Logic+? – continued

9.

The first screen allows you to set the Type of cabinet. The only

difference between the Type A1 and Type A2 configuration setting

is that the airflow velocity limits for normal operation of the
optional airflow sensor (because the Type A1 has a nominal inflow

of 85 FPM, and the Type A2 is 105 FPM). If the air velocity goes
out of range, the alarm will be displayed, but the cabinet will
continue to run. Selecting Type B2 will cause an inflow alarm if the

inflow velocity drops below the alarm setpoint, and will result in the
cabinet blower stopping. Use the

or

buttons to select the

cabinet Type, and then press

OK/Mute.


10.

The second screen allows you to configure the cabinet for a

canopy or not.*




11.

This screen configures the unit for a UV light or not. After your

selection is chosen, and you press

OK/Mute

, the unit will shut off

and then reset in the new configuration.





12.

If the unit is configured to NOT have a UV light, and the user

presses the UV Light button, they will get this message.







*Early production models of Type A Logic+ units may require a jumper wire

across the canopy electrical connection located on the left side of the electronics
module. Contact Labconco for assistance if this jumper is missing, and the unit

in constant canopy alarm mode.

Configuration

Type A Canopy Option

This cabinet has:

or

to choose,

OK to select, MENU to escape

a canopy

no canopy

 

This Logic is configured:

or

to choose,

OK to select, MENU to return

to have a UV light

to NOT have a UV light

 

UV Settings

This cabinet is not configured

for a UV light

If you want to install a UV

light, contact Labconco Product

Service at 800-821-5525

Attention

MENU to return