Labconco PURICARE Procedure Station Models 31260 User Manual

Page 90

Advertising
background image

Product Service 1-800-522-7658

90

Table 5-6

Secondary Inflow Test Specifications – 8 Inch Sash Opening

Cabinet Width (feet)

3

4

5

6

Sash Opening Template

1

3+

4+

5+

6+

Sensor distance (inches)

5

4

3.25

3.25

4

# of Test points

6

8

10

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (FPM)

213

283

353

231

Avg. Inflow Vel. Range (FPM)

203-223

269-296

336-370

220-242

Correction Factor (CF)

1

1

1

1.83

Average Inflow Volume Range (AIV) (Avg. velocity x CF)

203-223

269-296

336-370

403-443

Sash Open Area (Sq. Ft)

2.03

2.69

3.36

4.03

Inflow Velocity Range (AIV / Sash open area)

100-110

100-110

100-110

100-110

Secondary Inflow Test Specifications – 9 Inch Sash Opening

Cabinet Width (feet)

3

4

5

6

Sash Opening Template

1

3+

4+

5+

6+

Sensor distance (inches)

2

4

3.25

3.25

4

# of Test points

6

8

10

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (FPM)

240

318

397

260

Avg. Inflow Vel. Range (FPM)

228-251

303-333

378-416

248-272

Correction Factor (CF)

1

1

1

1.83

Average Inflow Volume Range (AIV) (Avg. velocity x CF)

228-251

303-333

378-416

453-498

Sash Open Area (Sq. Ft)

2.28

3.03

3.78

4.53

Inflow Velocity Range

(AIV / Sash open area)

105+/-5 105+/-5

105+/-5

105+/-5

Secondary Inflow Test Specifications – 10 Inch Sash Opening

Cabinet Width (feet)

3

4

5

6

Sash Opening Template

1

3+

4+

5+

6+

Sensor distance (inches)

5

4

3.25

3.25

4

# of Test points

6

8

10

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (FPM)

266

354

441

289

Avg. Inflow Vel. Range (FPM)

253-278

337-371

420-462

275-302

Correction Factor (CF)

1

1

1

1.83

Average Inflow Volume Range (AIV) (Avg. velocity x CF)

253-278

337-371

420-462

503-553

Sash Open Area (Sq. Ft)

2.53

3.37

4.2

5.03

Inflow Velocity Range (AIV / Sash open area)

100-110

100-110

100-110

100-110

Secondary Inflow Test Specifications – 12 Inch Sash Opening

Cabinet Width (feet)

3

4

5

6

Sash Opening Template

1

3+

4+

5+

6+

Sensor distance (inches)

5

4

3.25

3.25

4

# of Test points

6

8

10

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (FPM)

259

344

428

281

Avg. Inflow Vel. Range (FPM)

243-274

323-364

403-454

264-297

Correction Factor (CF)

1

1

1

1.83

Average Inflow Volume Range (AIV) (Avg. velocity x CF)

243-274

323-364

403-454

483-544

Sash Open Area (Sq. Ft)

3.04

4.04

5.04

6.04

Inflow Velocity Range (AIV / Sash open area)

85+/-5

85+/-5

85+/-5

85+/-5

Advertising