Section Two –

Installation Requirements

This section offers an overview of site requirements and how to prepare the

unit for operation.

Prerequisites

Before you install the Logic+, you need to prepare the site for installation.

Examine the location where you intend to install the cabinet. You must be

certain that the area is level and of solid construction. In addition, a

dedicated source of electrical power must be located near the installation

site.
Carefully read this chapter to learn:

Location requirements.

Electrical power requirements.

Exhaust requirements.

Service utility requirements.

Space requirements.

Refer to

Appendix C: Specifications

, for complete biosafety cabinet electrical

and environmental conditions, specifications and requirements

.

Space Requirements

The dimensions for the Logic+ biosafety cabinets are shown in

Section Nine:

Miscellaneous

.

Clearance

A minimum clearance of at least 6 inches (150 mm) is suggested on the top

and both sides of the cabinet for service.