Type a hepa filter leak test information, Adjusting the optional inflow and downflow values, If the logic+ is contaminated – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 72

Advertising
background image

Product Service 1-800-522-7658

72

Adjusting the optional inflow and
downflow values

Note: This step is only available if the optional airflow sensor is installed.

1.

Using the

or

buttons, increase or decrease the displayed inflow velocity

until it matches the observed value, then press “

OK/Mute

” to confirm the value.

The display will now change to downflow velocity calibration screen.

2.

Using the

or

buttons, increase or decrease the displayed downflow velocity

until it matches the observed value, then press “

OK/Mute

” to confirm the value.

3.

Once the downflow is confirmed, the unit will then automatically turn off and

then back on.

Type A HEPA Filter Leak Test

Information

All Logic+ models were tested to the HEPA Filter Leak Test as described in ANSI/NSF

Standard 49:2012.

Note: ensure all internal surfaces of the Logic+ have been appropriately

decontaminated before proceeding.

Figure 5-1

Note: For all Logic+ models, the aerosol

generator should be placed in the left rear

corner of the work area, pointing toward the

rear grille, as shown in Figure 5-1.


If the Logic+ is Contaminated

(An upstream concentration cannot be established)

If the Logic+ has been used with biohazardous, toxic chemicals, or radioisotopes, the

DOP access port cannot be used. The upstream concentration cannot be measured

directly. Testing at Labconco has shown the actual concentration varies from the

calculated value. Table 5-1 shows the actual vs. calculated concentrations for the Type A

models.