Miscellaneous, Troubleshooting – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 166

Advertising
background image

Product Service 1-800-522-7658

166

Section Nine –

Miscellaneous

Troubleshooting

Refer to the following table if the Biosafety Cabinet fails to operate properly. If the

suggested corrective actions do not solve the problem, contact Labconco for additional

assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Cabinet blower and

lights won’t turn on

Unit not plugged

into outlet

Plug the Biosafety Cabinet into

appropriate electrical service.

Check connection to control box on

top of Cabinet.

System Reset

Switch is Off

Turn on the System Reset Switch.

Circuit breaker(s)

tripped

Reset circuit breakers.

Keypad

disconnected or
defective

Run keypad diagnostics and check

connections.

Blower only runs at

low speed, but

lights work

Sash closed-unit is

in Night-Smart

Mode

Raise sash.

Sash Closure

Switch is

disconnected or

defective

Run keypad diagnostics and check

connections.

Motor signal wires

damaged or
disconnected

Measure motor output signal

voltage. Inspect signal wiring and
connections

Power Supply

Board

misconnected or

defective

Test output voltages on the Power

Supply board.