Labconco PURICARE Procedure Station Models 31260 User Manual

Page 146

Advertising
background image

Product Service 1-800-522-7658

146

Service Fixture Kit No. 3747500

Completely decontaminate all interior work surfaces of the Purifier before

beginning the installation of the service valve.

Additional service fixtures can be installed in any of the four available service fixture

holes in the sidewalls. The service fixture kit is Labconco part number 3747500.

Installation Procedure

1.

Decide which of the available locations you want to install the service fixture.

2.

Access the exterior surface of the sidewall by removing the two Phillips screws on

the front edge of the panel. Swing open and lift off the panel to access the valve

hole plug.

3.

With a flat blade screwdriver, carefully break the silicone seal on the perimeter of

the hole plug (on the outside of the cabinet). Then pry the plug away from the side

of the cabinet.


4.

Advertising