Quickstart, Main menu 1. mylogic, Settings 3. service – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 17: Service

Advertising
background image

Product Service 1-800-522-7658

17

Quickstart


How do I run a diagnostic on the Canopy Alarm?


1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.





4.

Press the

button until the

Diagnostic

option is highlighted (it

will turn white).

5.

Press

OK/Mute

to enter the Diagnostic submenu.





6.

Press the

button until the Diagnostic

Canopy Alarm

option is

highlighted (it will turn white). Press

OK/Mute

.






7.

The display shows the status of the Canopy Inlet Valve. During

normal operation, it will be open, and during a canopy alarm, it
will be closed. NOTE: The canopy door function is OPPOSITE OF

THE MAGNETIC SWITCH FUNCTION! – WHEN THE INLET VALVE
IS OPEN, THE MAGNETIC SWITCH IS CLOSED.








MAIN MENU

1. MyLogic

TM

2. Settings

3. Service

 

or

to choose,

OK to select, MENU to escape

or

to choose,

OK to select, MENU to return

Current Settings

Restore Default

Calibration
Configuration

Diagnostics

 

Service

Diagnostics

or

to choose,

OK to select, MENU to escape

Canopy Alarm

Sash sensors

Keypad - Relays

Airflow Sensor

 