Labconco PURICARE Procedure Station Models 31260 User Manual

Page 168

Advertising
background image

Product Service 1-800-522-7658

168

PROBLEM

CAUSE

CORRECTIVE ACTION

Fluorescent light

is dim or

flickering

Lamp(s) are

defective

Replace defective lamp(s)

Lamp wiring is

disconnected

Inspect lamp wiring.

Defective lamp

ballast

Replace lamp ballast.

UV light not

working

Sash is open

Close sash – UV light will not work

with the sash open.

Lamp is defective

Replace defective lamp.

Lamp wiring is

disconnected.

Inspect lamp wiring.

Defective lamp

ballast

Replace lamp ballast.

Keypad

disconnected or

defective

Run keypad diagnostics and check

connections.

UV light is dim or

flickering

Lamp is defective

or is at end of

operating lifetime.

Replace defective or worn out lamp.

Lamp wiring is

disconnected

Inspect lamp wiring.

Defective lamp

ballast

Replace lamp ballast.

Airflow Alert

goes off and/or

there is a slight
decrease in filter

life remaining

gauge

HEPA filter loading

The gauge reading steadily

decreases as the Cabinet is used.

Blockage of the

return air slots or

grille

Check all return air slots and grilles

to ensure that they are not blocked

or restricted.

Blockage of the

Ensure that the exhaust outlet is