Quickstart – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 11

Advertising
background image

Product Service 1-800-522-7658

11

Quickstart


How do I recalibrate the

blower speed

filter life gauge

airflow sensor reading

of the Logic+? - Continued

11.

If this is an initial certification or a recertification after HEPA filter

replacement, select

Reset Filter Gauge to 100%

. If this is an

annual recertification, select

Leave Gauge unchanged

. If you

want to set the filter gauge at a preset value, select

Set Gauge

to new setting

. Then press

OK/Mute

.




12.

The Type A units with airflow sensor will ask you to now

calibrate the average downflow to match the value you obtained
when you measured it. After the downflow is calibrated, and you

press

OK/Mute

, the Logic+ will power down, and then restart

with the new settings.


13.

If the unit is a Type A without an airflow sensor, after the filter

gauge option is selected and you press

OK/Mute

, the Logic+

will power down, and then restart with the new settings. If the

Unit is a Type B2 or a Type A with an airflow sensor, you will
then be asked to calibrate the inflow sensor to match the value

you obtained when you measured the inflow. After the inflow is
calibrated, and you press

OK/Mute

, the Logic+ will power

down, and then restart with the new settings.


14.

For the Type B2, the unit will now ask you to reduce the

total

exhaust volume to 80% of the nominal value. When you have
done this, press

OK/Mute

.

This step is critical-it sets the alarm

point for the exhaust flow. If you press OK/Mute before reducing
the flow to 80%, the Logic+ will assume the alarm setpoint is
equal to the nominal airflow, and the unit will return an Alarm-

Inflow Screen.

The Logic+ will power down, and then restart

with the new settings. Be sure to reset the exhaust damper to its
nominal setting.




Calibration

Filter Gauge Adjustment

or

to change filter gauge,

OK to select, MENU to finish

 

Reset Filter Gauge to 100%

Leave Gauge unchanged

Set Gauge to new setting

Reduce the exhaust flow to
80% of nominal volume. Press
OK when you are ready to fix
the setpoint.

Calibration

Inflow Alarm Setpoint

OK to select, MENU to finish

The Inflow displayed
value is: 102 FPM

Calibration

Inflow Display

or

to change display value,

UV Light to reset to 105 FPM,
OK to select, MENU to finish

 