Quickchart – logic+ type a1 - metric – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 22

Advertising
background image

Product Service 1-800-522-7658

22

QuickChart – Logic+ Type A1 - Metric

Model

30132

30142

30152

30162

Type

A1

A1

A1

A1

Cabinet Size (in meters)

.91

1.22

1.52

1.83

Sash Opening (inches)

305

305

305

305

Starting Serial #

1303_

1303_

1303_

1303_

Downflow Data

Nominal Avg. Downflow (m/s)

.279 +/-.03

.279 +/-.03

.279 +/-.03

.279 +/-.03

grid # of points (rows x columns)

21 (3x7)

24 (3x8)

30(3x10)

36(3x12)

Grid distance from back & sides

152

152

152

152

Distance between rows (mm)

141

141

141

141

Distance between columns (mm)

104

132

137

140

Inflow Data

Nominal Average Inflow (m/s)

.432 +/-.03

.432 +/-.03

.432 +/-.03

.432 +/-.03

Sash Open Area (Sq. m)

.283

.375

.468

.561

Nominal Avg. Exhaust Vol. (m3/s)

.122

.162

.202

.242

Avg. Exhaust Vol. Range (m3/s)

.115-.129

.153-.172

.190-.214

.228-.257

Secondary Inflow Data

3

Sash Opening Template

1

3+

4+

5+

6+

Sensor distance (mm)

2

102

83

83

102

# of Test points

6

8

10

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (m/s)

1.32

1.75

2.17

1.43

Avg. Inflow Vel. Range (m/s)

1.23-1.39

1.64-1.85

2.05-2.31

1.34-1.51

Correction Factor (CF)

.0929

.0929

.0929

.1700

Average Inflow Volume Range

(AIV) (Avg. velocity x CF)

.115-.129

.153-.172

.190-.214

.228-.257

Sash Open Area (Sq. m)

.283

.375

.468

.561

Inflow Velocity Range

(AIV / Sash open area)

.432 +/-.03

.432 +/-.03

.432 +/-.03

.432 +/-.03

HEPA Filter Leak Test Data

Air Displacement (m/s)

.283

.378

.472

.566

Laskin Nozzles needed

1

2

2

2

Theoretical aerosol conc. (ug/l)

3

23

34

27

23

Actual aerosol conc. (ug/l)

4

12

18

15

15

Supply HEPA Data

Width x Depth x Height (in.)

5

36x18x3.06

48x18x3.06

60x18x3.06

72x18x3.06

Performance (CFM)

335

445

555

665

Performance (Pressure in. H

2

O)

< .60”

<.45”

<.45”

<.42”

Labconco P/N

3838400

3838401

3838402

3838403

Exhaust HEPA Data

Width x Depth x Height (in.)

5

18x18x5.88

26x18x5.88

36x18x5.88

48x18x5.88

Performance (CFM)

266

353

441

529

Performance (Pressure in. H

2

O)

< 0.35”

< 0.28”

< 0.28”

< 0.28”

Labconco P/N

3838500

3838501

3838502

3838503

Motor/Blower Data

Labconco P/N

6

3832200

3832201

3832207

3832208

Motor HP

½

½

¾

¾

Fluorescent Lamp Data

Fluorescent Lamps (2 each)

F25T8 TL741 F32T8 TL741

F40T8 SP41

F40T8 SP41

Fluor. Lamp Labconco Part #

9721901

9721900

9721903

9721903

Fluor. Ballast Labconco Part #

3838100

3838100

3838100

3838100

Ventus Canopy Data

Labconco Canopy P/N

3889200

3889201

3889202

3889203

Canopy Slot Area (m

2

)

.0111

.0111

.0111

.0111

Canopy Inflow Range (m/s)

.010-.047

.010-.047

.010-.047

.010-.047

Nominal Canopy Slot Velocity (m/s)

1.22-1.32

1.22-1.32

1.22-1.32

1.22-1.32

Nominal Canopy Slot Volume (m3/s)

.235

.235

.235

.235

Canopy Vacuum (Pa)

25-37

25-37

25-37

25-37

Advertising