Quickchart – logic+ type a1 - imperial units – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 21

Advertising
background image

Product Service 1-800-522-7658

21

QuickChart – Logic+ Type A1 - Imperial Units

Model

30132

30142

30152

30162

Type

A1

A1

A1

A1

Cabinet Size (in feet)

3

4

5

6

Sash Opening (inches)

12

12

12

12

Starting Serial #

1303_

1303_

1303_

1303_

Downflow Data

Nominal Avg. Downflow (FPM)

55+/-5

55+/-5

55+/-5

55+/-5

grid # of points (rows x columns)

21 (3x7)

24 (3x8)

30(3x10) 36(3x12)

Grid distance from back & sides

6.0

6.0

6.0

6.0

Distance between rows

5.55

5.55

5.55

5.55

Distance between columns

4.08

5.21

5.39

5.50

Inflow Data

Nominal Average Inflow (FPM)

85+/-5

85+/-5

85+/-5

85+/-5

Sash Open Area (Sq. Ft)

3.04

4.04

5.04

6.04

Nominal Avg. Exhaust Vol. (CFM)

259

344

428

514

Avg. Exhaust Vol. Range (CFM)

243-274

323-364

403-454

483-544

Secondary Inflow Data

3

Sash Opening Template

1

3+

4+

5+

6+

Sensor distance (inches)

2

4

3.25

3.25

4

# of Test points

6

8

10

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (FPM)

259

344

428

281

Avg. Inflow Vel. Range (FPM)

243-274

323-364

403-454

264-297

Correction Factor (CF)

1

1

1

1.83

Average Inflow Volume Range

(AIV) (Avg. velocity x CF)

243-274

323-364

403-454

483-544

Sash Open Area (Sq. Ft)

3.04

4.04

5.04

6.04

Inflow Velocity Range

(AIV / Sash open area)

85+/-5

85+/-5

85+/-5

85+/-5

HEPA Filter Leak Test Data

Air Displacement (CFM)

600

800

1000

1200

Laskin Nozzles needed

1

2

2

2

Theoretical aerosol conc. (ug/l)

3

23

34

27

23

Actual aerosol conc. (ug/l)

4

12

18

15

15

Supply HEPA Data

Width x Depth x Height (in.)

5

36x18x3.06

48x18x3.06

60x18x3.06

72x18x3.06

Performance (CFM)

335

445

555

665

Performance (Pressure in. H

2

O)

< .60”

<.45”

<.45”

<.42”

Labconco P/N

3838400

3838401

3838402

3838403

Exhaust HEPA Data

Width x Depth x Height (in.)

5

18x18x5.88

26x18x5.88

36x18x5.88

48x18x5.88

Performance (CFM)

266

353

441

529

Performance (Pressure in. H

2

O)

< 0.35”

< 0.28”

< 0.28”

< 0.28”

Labconco P/N

3838500

3838501

3838502

3838503

Motor/Blower Data

Labconco P/N

6

3832200

3832201

3832207

3832208

Motor HP

½

½

¾

¾

Fluorescent Lamp Data

Fluorescent Lamps (2 each)

F25T8

TL741

F32T8

TL741

F40T8

SP41

F40T8

SP41

Fluor. Lamp Labconco Part #

9721901

9721900

9721903

9721903

Fluor. Ballast Labconco Part #

3838100

3838100

3838100

3838100

Ventus Canopy Data

Labconco Canopy P/N

3889200

3889201

3889202

3889203

Canopy Slot Area (F

2

)

.12

.12

.12

.12

Canopy Inflow Range (CFM)

20-100

20-100

20-100

20-100

Nominal Canopy Slot Velocity (FPM)

240-260

240-260

240-260

240-260

Nominal Canopy Slot Volume (CFM)

50

50

50

50

Canopy Vacuum (In. H

2

O)

0.1-0.15

0.1-0.15

0.1-0.15

0.1-0.15